top of page

Meggitt  Piezoelectric (charge mode)  accelerometer

Piezoelectic 가속도계는 charge mode 출력형입니다.  외부 charge amplifier 또는 charge converter와 접속하여 사용합니다. 

넓은 동작 온도와 넓은 진폭 범위에서 안정적으로 동작하며,  고온 +815℃ (+1500℉), 방사능 환경에서 사용되는 특수 목적의 센서 모델도 준비되어 있습니다.

  Meggitt의 압전 가속도계는 원자력 발전소, 제철소 및  광범위한 제트 엔진 (민간 및 군용), 헬리콥터 및 우주 응용 분야에도 이상적으로   적합합니다.

  • gas turbine engine monitoring

  • airframe structure vibration analysis

  • gearbox analysis

  • bearing analysis

  • Health and usage monitoring systems (HUMS)

  • rotor trim and balance

  • propeller balancing

  • auxiliary power unit monitoring

  • cryogenic environments on space applications

temp1.png

+260 ℃ / +399  ℃

2273A

2273AM1 / 2273AM20

Meggitt 메기트 가속도 센서2273A.jpg

▣ Key features
   • Radiation hardened
   • Shock limit 10 000 gpk
   • Case grounded
   • Operates over wide temperature range
   • Vibration measurements in nuclear, and high
     temperature environments

Specifications
   • Sensitivity: 3pC/g
   • Amplitude response: 1~6,000Hz ±1dB
   • Temperature range: -184~+399℃ (-300~+750℉)
   • Output Resistance at +399℃: ≥10MΩ
   • Integrated gamma flux: up to 6.2 x 10^10 [rad]

   • Integrated neutron flux: up to 3.7 x 10^18 [N/cm^2]
   • Weight: 25gram
   • Case material: Stainless steel
▣ Applications
   • Test cell vibration measurements
   • Nuclear and high temperature applications

2273AM.jpg
▣ Key features
   • High temperature operation (+399°C)
   • Radiation-hardened
▣ Applications

   • Sensitivity: 10pC/g 
   • Amplitude response: 1~7,000Hz ±1dB
   • Temperature range: -55~+399℃ (-67~+750℉)
   • Resistance min at 399℃: ≥10MΩ
   • Signal ground isolated
   • Weight: 33gram
   • Case material: Inconel

▣ Applications
   • Test cell vibration measurements
   • Reactor and loose parts testing

6222M59

6222M59.jpg

▣ Key features
   • Rugged hermetic design
   • Balanced differential output
   • Ground isolated

Specifications
   • Sensitivity: 50pC/g
   • Charge sesitivity deviation vs frequency:
                                  ±5% max. 20Hz to 4kHz ±5%

   • Temperature range: -53~+260℃ (-65~+500℉)
   • Internal Resistance at 100Vda: 50MΩ min at 450F(232℃)
   •• Sinusoidal limit: 200g, Shock limit: 1,000g

   • Case material: Stainless steel
▣ Applications
   • Gas-turbine monitoring

temp2.png

+482 ℃ 

2278-1

2278-1.jpg
▣ Key features
    • High temperature operation +1200ºF(+650ºC)
    • Ground Isolated
    • Small and lightweight, 1.7 ounces (48 gm)

    • 1축형
▣ Specifications
    • Sensitivity: 4pC/g
    • Frequency response: 20~4,000Hz ±5%
    • Temperature range: -55~+650℃ (-67~+1200℉)
    • Resistance min at max temp: ≥10kΩ
    • Signal ground isolated,
    • Case material: Inconel
    • Weight: 48gram
◈ Applications
    • Gas turbine testing
    • ​Nuclear applications

2278

2278.jpg
2278.PNG
▣ Key features
   • 3축형
   • Ground Isolated
   • Small and lightweight, 65 grams 


▣ Specifications
   • Sensitivity: 4pC/g 
   • Amplitude response:  Z축 10~6,000Hz ±10% 
                                   X축 & Y축 10~800Hz ±10% 

   • Temperature range: -55~+650℃ (-67~+1200℉)
   • Resistance min at max temp: ≥10kΩ
   • Signal ground isolated
   • Weight: 65gram


▣ Applications
   • Gas Turbine testing
   • Nuclear applications
2280.jpg
▣ Key features
   • 3축형
   • High temperature operation 900°F (+482ºC)
   • Ground Isolated
   • Rugged design
▣ Specifications
   • Sensitivity: 3pC/g, 

   • Frequency response: 10~4,000Hz ±5%
   • Temperature range: -54~+482℃ (-65~+900℉)
   • Resistance min at max temp: ≥100kΩ
   • Signal ground isolated
   • Case material: Inconel
   • Weight: 250gram

▣ Applications
   • Gas Turbine testing
   • Nuclear applications
2276.jpg
▣ Key features
   • Radiation hardened
   • Inconel construction
   • Requires no external power
▣ Specifications

   • Sensitivity: 10pC/g, 
   • Frequency response: 1~7,000Hz  ±1dB
   • Temperature range: -55~+482℃ (-67~+900℉)
   • Resistance min at max temp: ≥100kΩ
   • Case grounded
   • Integrated gamma flux: up to 6.2 x 10^10 [rad]
   • Integrated neutron flux: up to 3.7 x 1018 [N/cm^2]
   • Case material: Inconel
   • Weight: 30gram

▣ Applications
   • Test cell vibration measurements
   • Nuclear and high temperature applications

2280

2276

2248.jpg

2248 / 2248M1

6233C

▣ Key features
   • 1축형
   • Small size
   • Light weight
▣ Specifications

   • Sensitivity: 3pC/g
   • Frequency response: 1~8,000Hz ±1dB
   • Temperature range: -54~+482℃ (-65~+900℉)
   • Resistance min at max temp: ≥100kΩ
   • Case grounded
   • Case material: Inconel

   • Weight: 13gram

   • Dimensions: 11.09 x 11.09 x 16.51
▣ Applications
   • Gas Turbine engine monitoring
   • Nuclear applications
6233C.png
6233.PNG
▣ Key features
   • 1축형
   • ARINC mount (3x 8-32 socket HD screw)
   • +900°F (+482°C) operation
  • Ground isolated
   • Balanced differential output

▣ Specifications
   • Sensitivity: 10/50/100pC/g,
   • Amplitude response: 1~10kHz / 5kHz / 4.5kHz (±1dB)
   • Temperature range: -55~+482℃ (-67~+900℉)
   • Resistance min at max temp: ≥100kΩ
   • Weight: 75 / 110 / 110gm
▣ Applications
   • Gas turbine monitoring
   • Test cell vibration measurements
   • Nuclear applications

6235M1

6240M4X / 5X / 6X

6235M1.jpg
▣ Key features
   • Requires no external power
   • +900°F (+482°C) operation
   • Hermetically sealed
   • Ground isolated
   • Balanced differential output
▣ Specifications

   • Sensitivity: 10pC/g
   • Sinusoidal limit: 1,000g, Shock limit: 2,000g
   • Amplitude response: 1~9,000Hz  ±1dB
   • Temperature range: -196~+482℃ (-325~+900℉)
   • Resistance min at max temp: 100kΩ
   • Weight: 75gram
   • Cable: Integral, Cable optional: 3090C-XXX
▣ Applications
   • Aircraft and gas turbine engine monitoring
   • Test cell vibration measurements
   • Nuclear applications
6240M456X.jpg
▣ Key features
   • Rugged hermetic design
• Balanced differential output
• Ground Isolated
• Various cable lengths and connectors

▣ Specifications
   • Sensitivity: 50pC/g,
   • Sinusoidal limit: 200g, Shock limit: 1,000g
   • frequency response: 20Hz~350Hz  ±5%
   • Temperature range: -54~+482℃ (-65~+500℉)
   • Resistance Pin to Pin at +482℃: 10kΩ
   • Weight: 30gram
   • Material: Inconel600
   • Cable: Hardline 
▣ Applications
  • Gas-turbine monitoring
temp3.png

+650℃ / +815℃

Extreme High Temp. Piezoelectric Accelerometer

6240M10

6243M8 / M9

6240M10.jpg
▣ Key features
   • +1200˚F (+650˚C) operation,
     +1400˚F (+760˚C) intermittent
   • Hermetically sealed
   • No pyroelectric or thermal velocity spiking
   • Single bolt mount (1/4-28 socket HD screw)
   • Ground isolated
▣ Specifications
   • Sensitivity: 5pC/g,

   • Sinusoidal limit: 250g, Shock limit: 1,000g
   • Amplitude response: 1~3,000Hz  ±1dB
   • Temperature range: -54~+650℃ (-65~+1,200℉)
   • Resistance min at max temp: ≥10kΩ
   • Weight: 95gram
   • Cable: Integral, Cable optional: 3090C-XXX
▣ Applications
   • Aircraft and gas turbine engine monitoring
   • Test cell vibration measurements
   • Nuclear applications
6240M8-9.jpg
6240M8-9.jpg
▣ Key features
   • +1000˚F (+538˚C) operation
   • High temperature softline cable provides easy cable
       routing in cramped locations
   • Hermetically sealed (accelerometer)
   • No pyroelectric or thermal velocity spiking
   • Single bolt mount
   • Ground isolated

▣ Specifications
   • Sensitivity: 5.5pC/g,
   • Sinusoidal limit: 500g, Shock limit: 2,000g
   • Amplitude response: 1~3,000Hz  ±10%
   • Temperature range: -54~+650℃ (-65~+1,200℉)
   • Resistance at over temp. range: ≥200kΩ
   • Weight: 30gram
   • Material: Inconel
   • Cable optional: 3090C-XXX
▣ Applications
   • Aircraft and gas turbine engine monitoring
   • Test cell vibration measurements
   • Nuclear applications
6243M 1 2.jpg
▣ Key features
   • +1200˚F (+650˚C) operation
   • Integral hardline cable
   • Hermetically sealed
   • No pyroelectric or thermal velocity spiking
   • Single bolt mount
   • Ground isolated

▣ Specifications
   • Sensitivity: 5.5pC/g,
   • Sinusoidal limit: 500g, Shock limit: 2,000g
   • Amplitude response: 1~3,000Hz  ±10%
   • Temperature range: -54~+650℃ (-65~+1,200℉)
   • Resistance Pin to Pin at +650℃: ≥10kΩ
   • Weight: 30gram
   • Material: Inconel
   • Cable optional: 3090C-XXX
▣ Applications
   • Aircraft and gas turbine engine monitoring
   • Test cell vibration measurements
   • Nuclear applications
6243M 3 4.jpg
▣ Key features
• +1200˚F (+650˚C) operation
• Integral hardline cable
• Hermetically sealed
• No pyroelectric or thermal velocity spiking
• Single bolt mount
• Balanced differential output

▣ Specifications
   • Sensitivity: 5.5pC/g,
   • Sinusoidal limit: 500g, Shock limit: 2,000g
   • Amplitude response: 1~3,000Hz  ±10%
   • Temperature range: -54~+650℃ (-65~+1,200℉)
   • Resistance Pin to Pin at +650℃: ≥10kΩ
   • Weight: 30gram
   • Material: Inconel
   • Cable: Hardline, two-pin connector 
▣ Applications
   • Aircraft and gas turbine engine monitoring
   • Test cell vibration measurements
   • Nuclear applications

6243M1 / 6243M2

6243M3 / 6243M4

6237M70 /M71

6245

6237M 70 71.jpg
▣ Key features
   • Integral hardline cable
   • Single bolt mount
   • Ground Isolated

▣ Specifications
   • Sensitivity: 10pC/g,
   • Sinusoidal limit: 500g, Shock limit: 2,000g
   • Frequency response: 1~5,000Hz  ±1dB
   • Temperature range: -54~+650℃ (-65~+1,200℉)
   • Resistance min at max temp: ≥10kΩ
   • Signal ground isolated, Weight: 30gram
   • Cable: Integral
   • Cable optional: 1001-ZZZ for under 288℃ (+550F)
   • Remote charge converter: 1772-2
▣ Applications
   • Gas Turbine Testing
   • Test cell
6245.jpg
▣ Key features
   • +1500˚F (+815˚C) operation
   • Integral hardline cable
   • Hermetically sealed
   • No pyroelectric or thermal velocity spiking
   • Single bolt mount
   • Ground isolated

▣ Specifications
   • Sensitivity: 3pC/g,

   • Sinusoidal limit: 500g, Shock limit: 2,000g
   • Frequency response: 1~4,000Hz  ±10%
   • Temperature range: -55~+815℃ (-65~+1,500℉)
   • Resistance min at max temp: ≥10kΩ
   • Weight: 30gram
▣ Applications
   • Aircraft and gas turbine engine monitoring
   • Gas turbine vibration measurements
   • Nuclear applications

Meggitt 고온용 가속도 / 압력 센서 판매점 : (주)주원 

경기도 성남시 분당구 성남대로43번길 10, 607호 

Tel. 031-726-1585

   M. 010-5640-1586

sbkim@joowonprecision.com  

bottom of page